Biomechanical analysis of two-point asymmetric screw fixation with implant for femoral neck fracture

Show simple item record

dc.contributor.author Karev, D. B.
dc.contributor.author Barsukov, V. G
dc.date.accessioned 2016-09-06T06:40:54Z
dc.date.available 2016-09-06T06:40:54Z
dc.date.issued 2012-04
dc.identifier.citation Karev, D. B. Biomechanical analysis of two-point asymmetric screw fixation with implant for femoral neck fracture / D. B. Karev,V. G. Barsukov // Acta mechanica et automatica. – 2012. – Vol. 6, № 4. – Р. 19-22. ru_RU
dc.identifier.uri http://elib.grsmu.by/handle/files/689
dc.description Femoral Neck Fracture, Screw Fixation, Stressed State, Cortical Bone, Trabecular Bone ru_RU
dc.description.abstract Stressed state peculiarities of cortical and trabecular bones by two-point asymmetric screw fixation with implant for femoral neck fracture are studied. Layer construction mechanic methods are used for analysis of stresses in cortical and trabecular bones. Biomechani-cal conditions for non-opening of the junction of the bone parts being joined are determined. It has been found that the total tightness of the broken parts when they rest against each other is secured over the whole fracture section without junction opening under condition that fixing screws are positioned in the trabecular bone without penetration of the thread side surface into cortical bone. ru_RU
dc.language.iso en ru_RU
dc.title Biomechanical analysis of two-point asymmetric screw fixation with implant for femoral neck fracture ru_RU